TG 50 Thermoplastic

Technical Drawing

RITTER TG50 Thermoplastic schematic

Legend

  1. Handle
  2. Support unit for:
    • Thermometer (gas)
    • Thermometer (packing liquid)
    • Manometer
  3. Level
  4. Removable cover
  5. Pulse generator port (option)
  6. Manometer connection
  7. Filling nozzle
  8. Rear plate, detachable
  9. Gas inlet
  10. Gas outlet
  11. Liquid level indicator
  12. Drainage faucet
  13. Levelling foot

Performance Data

Minimum flow Qmin100ltr/h
Standard flow Qstandard10,000ltr/h
Maximum flow Qmax18,000ltr/h
Measurement accuracy at standard flow± 0.2%
Measurement accuracy across measuring range± 0.5%
Measuring drum volume50ltr/U
Packing liquid quantity, approx91ltr
Maximum gas inlet pressure50mbar
Minimum differential pressure (1)0.1mbar
Minimum dial division 0.5ltr
Maximum indication value (2)999,999,990ltr
Connection gas in/outlethose barb
Inside-Ø hose barb31.8mm
Outside-Ø hose barb40mm

(1) Differential pressure (= pressure loss) gas inlet gas outlet
(2) Standard Totalizing Roller Counter

Models (materials)

Model Casing Measuring drum Width Height Depth Weight
(without packing liquid)
Max. constant
use temperature
5 PVC-transparent PVC-grey 700 mm 725 mm 755 mm 57.0 kg 40 °C
6 PP-grey PP-grey 700 mm 725 mm 755 mm 32.0 kg 60 °C
7 PVDF PVDF 700 mm 725 mm 755 mm 73.3 kg 80 °C
8 PE-el PE-el 700 mm 725 mm 755 mm 40.7 kg 80 °C

Caution: Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion! For chemical resistance properties please contact your RITTER distribution partner!

Standard Equipment

  • Multi-Chamber Measuring Drum
  • Magnetic coupling between measuring drum and counter (for an absolutely liquid-tight casing)
  • 8 or 9-digit Totalizing Roller Counter
  • Large One-needle Dial
  • Supports for Thermometer and Manometer
  • Viton sealing
  • Level and Levelling Feet
  • Bubble level for levelling

Built-in Options

Accessories

V 2.1 / Rev. 2024-07-03 / Subject to alterations.