Analog Voltage Output

RITTER Data sheet
Analog Voltage Output

Overview

The Analog Voltage Output Module allows the simultaneous output of up to 4 different values at 4 separate channels:

  • 4 parallel signal outputs
  • The measured values of the oxygen and humidity sensor can only be transmitted alternatively
  • Resolution: 16 bit
  • Update rate: 1 sec

Technical Data

For connection to an analog data acquisition device the analog voltage output port alternatively provides the following voltage levels:

  • 0 – 2 V
  • 0 – 5 V
  • 0 – 10 V

The voltage range is pre-set according to the order and cannot be changed by the user.

Please note the assignments of the respective gas concentration measurements to the respective channels. These assignments are stated in the Calibration Protocol of the sensor unit.

Assignment of contacts to the channels of the gas concentration measurements at the socket and plug of the Analog Voltage Output:

image003
View to socket
image002
View to plug
Contact No.Assignment to channelWire Colors of provided Connection Cable
1channel 1white
2channel 2green
3channel 3yellow
4channel 4grey
5groundbrown

Analogue output signals of oxygen and humidity sensors:

  • NDIR sensors: Channel 4 (grey wire)
  • NDUV sensors: Channel 2 (green wire)

Only one of these two signals can be transmitted alternatively.

The connection cable is delivered with open wires to be connected to the analog data acquisition device of the user. The wire colours are listed in the above table.

V 1.0 / Rev. 2025-09-04 / Subject to alterations.