MultiGas 红外传感器 NDIR

CO2 / CO / NO / CH4 / CnHm / N2O
  • NDIR传感器=非色散红外传感器。
  • 采用高性能发光二极管(IR-LED)和热微辐射器。

概述

CO2 CO N2O CH4 CnHm* CF4 SF6 H2O

The RITTER MultiGas Sensors NDIR Module (non-dispersive IR sensor) has been specially developed for use in high-quality gas analysis. In the design phase special emphasis was placed on high stability and a low detection limit. These goals could be fully achieved by using high-performance light-emitting diodes (IR-LED) and thermal micro-radiators which were adapted to the requirements of gas detection. In the spectral range from 2 μm to 12 μm, carbon dioxide, carbon monoxide, hydrocarbons, water vapor and sulfur hexafluoride can be safely detected down to the ppm range with this innovative sensor platform.

The various photometric components such as detectors, emitters, measuring sample cell, etc. will be assembled user-specifically in a high-quality tabletop casing by RITTER.
RITTER MultiGas Sensorgehäuse

应用程序

  • Biogas-Forschung
  • Umwelt- und Prozessmesstechnik
  • Elementaranalyse
  • TOC-Analysatoren
  • Industrielle Gasanalyse
  • Erdgasanalyse

Characteristics and Benefits

  • Group of detectable gases:
    CO2 CO N2O CH4 CnHm* CF4 SF6 H2O
  • Measurement technology: NDIR-Sensor (non-dispersive IR sensor)
  • 测量精度 ±2% of Full Scale (F.S.)
  • No cross-sensitivity to H2
  • Operating temperature: 5 – 45 °C
  • Operating pressure: 800 – 1200 mbar (hPa) abs.
  • Flow rate range: 1 ltr/day – 100 ltr/h
  • Warm-up time: 2 min
  • Response time (t90): ≈ 3 sec
  • Interface USB, on request RS 232
  • Sensor cuvette: Aluminum, gold plated for sensor length 100 ≥ mm
  • Incl. temperature compensation
  • Incl. data acquisition software
  • In tabletop casing, overall dimensions W x H x L 171 x 85 x 246 mm, weight approx. 1,9 kg
  • Gas connection: PVDF screw-type tube connection for tube Øi 4 mm, Øa 6 mm
  • Power supply: 24 VDC (incl. power plug 100 – 240 VAC / 24 VDC)

Recalibration

  • Readjustment of the zero-point is recommended once a week.
  • Contamination of the measuring cuvette can lead to shifting of the end-point. Due to the integrated filter, re-adjustment of the end-point is generally not required – but can be performed once a year.
  • Additionally, cartridges filled with an inert and the respective test gas for recalibration of the zero and end-point can be provided.
  • Long-term drift is < 1% of Full Scale (F.S.) / 24 h.

Options / Additional Sensors

氧气传感器

Overview The oxygen sensor is …

压力传感器

Overview A change in gas press …

湿度传感器

The humidity sensor is a senso …

模拟电压输出

概述 为了连接到模拟数据采集配置,模拟电压输出端口(位于传感 …