TG 3 Stainless steel

tg003.0 14 v33

Performance Data

Minimum flow Qmin 6 ltr/h
Standard flow Qstandard 300 ltr/h
Maximum flow Qmax 360 ltr/h
Measurement accuracy at standard flow 0.2 %
Measurement accuracy across measuring range 0.5 %
Measuring drum volume 3 ltr/U
Packing liquid quantity, approx 11 ltr
Maximum gas inlet pressure (1) 500 mbar
Minimum differential pressure (2) 0.2 mbar
Minimum dial division 0.02 ltr
Maximum indication value (3) 99,999,999 ltr
Connection gas in/outlet
Inside-Ø 10.6 mm
Outside-Ø 14 mm
  • (1) Standard version: 500 mbar – special version also available for 1 bar
  • (2) Differential pressure (= pressure loss) gas inlet gas outlet
  • (3) Standard Totalizing Roller Counter

Models (materials)

Model Casing Measuring drum Width Height Depth Weight
(without packing liquid)
Max. constant
use temperature
1 Stainless Steel PVC-grey 363 mm 410 mm 445 mm 15.8 kg 40 °C
2 Stainless Steel PE-el 363 mm 410 mm 445 mm 15.7 kg 60 °C
3 Stainless Steel PP-grey 363 mm 410 mm 445 mm 15.7 kg 80 °C
4 Stainless Steel PVDF 363 mm 410 mm 445 mm 16.2 kg 80 °C

Caution

Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion!

For chemical resistance properties please contact your RITTER distribution partner!

Standard Equipment

  • 8-digit Totalizing Roller Counter
  • Bubble level for levelling
  • Level and Levelling Feet
  • Magnetic Coupling (between the measuring drum and counting mechanism)
  • Multi-Chamber Measuring Drum
  • Supports for Thermometer and Manometer
  • Viton sealing

V 3.3 / Rev. 2022-06-27 / Subject to alterations.