tg003.0 58 v21

Legend

Performance Data

Minimum flow Qmin 6 ltr/h
Standard flow Qstandard 300 ltr/h
Maximum flow Qmax 360 ltr/h
Measurement accuracy at standard flow ± 0.2 %
Measurement accuracy across measuring range ± 0.5 %
Measuring drum volume 3 ltr/U
Packing liquid quantity, approx 5.8 ltr
Maximum gas inlet pressure (1) 50 mbar
Minimum differential pressure (2) 0.2 mbar
Minimum dial division 0.02 ltr
Maximum indication value (3) 99,999,999 ltr
Connection gas in/outlet
Inside-Ø 9.5 mm
Outside-Ø 16 mm
  • (1) Standard version: 50 mbar – special versions up to 500 mbar available
  • (2) Differential pressure (= pressure loss) gas inlet gas outlet
  • (3) Standard Totalizing Roller Counter

Models (materials)

Model Casing Measuring drum Width Height Depth Weight
(without packing liquid)
Max. constant
use temperature
5 PVC-transparent PVC-grey 330 mm 375 mm 405 mm 6.3 kg 40 °C
6 PP-grey PP-grey 330 mm 375 mm 405 mm 4.5 kg 80 °C
7 PVDF PVDF 330 mm 375 mm 405 mm 8.1 kg 80 °C
8 PE-el PE-el 330 mm 375 mm 405 mm 4.5 kg 60 °C

Caution

Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion! For chemical resistance properties please contact your RITTER distribution partner!

Standard Equipment

  • 8-digit Totalizing Roller Counter
  • Bubble level for levelling
  • Level and Levelling Feet
  • Magnetic Coupling (between the measuring drum and counting mechanism)
  • Multi-Chamber Measuring Drum
  • Supports for Thermometer and Manometer
  • Viton sealing

V 2.1 / Rev. 2022-06-27 / Subject to alterations.