TG 1 Stainless steel 6 bar

TG1 40bar

Legend

  1. Level
  2. Levelling foot
  3. Handle
  4. Support unit for:
    • Thermometer gas (electric)
    • Thermometer packing liquid (electric)
  5. Removable cover
  6. Pulse generator port (option)
  7. Gas outlet
  8. Filling nozzle
  9. Manometer connection
  10. Level indicator
  11. Gas inlet
  12. Drainage faucet
  13. Levelling foot

Performance Data

Minimum flow Qmin 2 ltr/h
Standard flow Qstandard 100 ltr/h
Maximum flow Qmax 120 ltr/h
Measurement accuracy at standard flow 0.2 %
Measurement accuracy across measuring range 0.5 %
Measuring drum volume 1 ltr/U
Packing liquid quantity, approx 4 ltr
Maximum gas inlet pressure 6 bar
Minimum differential pressure (1) 0.2 mbar
Minimum dial division 0.01 ltr
Maximum indication value (2) 99,999,999 ltr
Connection gas in/outlet hose barb
Inside-Ø hose barb 10.6 mm
Outside-Ø hose barb 12 mm
(1) Differential pressure (= pressure loss) gas inlet gas outlet
(2) Standard Totalizing Roller Counter

Models (materials)

Model Casing Measuring drum Width Height Depth Weight
(without packing liquid)
Max. constant
use temperature
1 Stainless Steel PVC-grey 270 mm 320 mm 400 mm 14.0 kg 40 °C
2 Stainless Steel PE-el 270 mm 320 mm 400 mm 13.9 kg 60 °C
3 Stainless Steel PP-grey 270 mm 320 mm 400 mm 13.9 kg 80 °C
4 Stainless Steel PVDF 270 mm 320 mm 400 mm 14.2 kg 80 °C

Caution

Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion!

For chemical resistance properties please contact your RITTER distribution partner!

Standard Equipment

  • Multi-Chamber Measuring Drum
  • Magnetic coupling between measuring drum and counter (for an absolutely liquid-tight casing)
  • 8 or 9-digit Totalizing Roller Counter
  • Large One-needle Dial
  • Supports for Thermometer and Manometer
  • Viton sealing
  • Level and Levelling Feet
  • Bubble level for levelling

Built-in Options

  • LCD Display, resettable, 8-digit (replaces Totalizing Roller Counter)
  • Pulse Generator, standard or Ex-proof version (for connecting Electronic Display Unit/Computer)

Accessories

V 1.2 / Rev. 2024-04-17 / Subject to alterations.