TG 0.5 Stainless steel 6 bar

TG05 Datenblatt 1 Dimensionen 1

Legend

  1. Handle
  2. Support unit for:
    • Thermometer (gas)
    • Thermometer (packing liquid)
  3. Level
  4. Removable cover
  5. Pulse generator port (option)
  6. Gas inlet
  7. Filling nozzle
  8. Levelling foot
  9. Gas outlet
  10. Manometer connection
  11. Liquid level indicator
  12. Drainage faucet

Performance Data

Minimum flow Qmin 1 ltr/h
Standard flow Qstandard 50 ltr/h
Maximum flow Qmax 60 ltr/h
Measurement accuracy at standard flow 0.2 %
Measurement accuracy across measuring range 0.5 %
Measuring drum volume 0.5 ltr/U
Packing liquid quantity, approx 2.7 ltr
Maximum gas inlet pressure 6 bar
Minimum differential pressure (1) 0.2 mbar
Minimum dial division 0.002 ltr
Maximum indication value (2) 9,999,999.9 ltr
Connection gas in/outlet hose barb
Inside-Ø hose barb 10.6 mm
Outside-Ø hose barb 15 mm
(1) Differential pressure (= pressure loss) gas inlet gas outlet
(2) Standard Totalizing Roller Counter

Models (materials)

Model Casing Measuring drum Width Height Depth Weight
(without packing liquid)
Max. constant
use temperature
1 Stainless Steel PVC-grey 300 mm 350 mm 305 mm 13.3 kg 40 °C
2 Stainless Steel PE-el 300 mm 350 mm 305 mm 13.2 kg 60 °C
3 Stainless Steel PP-grey 300 mm 350 mm 305 mm 13.2 kg 60 °C
4 Stainless Steel PVDF 300 mm 350 mm 305 mm 13.5 kg 60 °C

Caution

Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion!

For chemical resistance properties please contact your RITTER distribution partner!

Standard Equipment

  • Multi-Chamber Measuring Drum
  • Magnetic coupling between measuring drum and counter (for an absolutely liquid-tight casing)
  • 8 or 9-digit Totalizing Roller Counter
  • Large One-needle Dial
  • Supports for Thermometer and Manometer
  • Viton sealing
  • Level and Levelling Feet
  • Bubble level for levelling

Built-in Options

  • LCD Display, resettable, 8-digit (replaces Totalizing Roller Counter)
  • Pulse Generator, standard or Ex-proof version (for connecting Electronic Display Unit/Computer)

Accessories

  • Software »RIGAMO« RIGAMO
  • Resistance Temperature Sensor (Gas), 100 mm
  • Resistance Temperature Sensor (Packing Liquid), 200 mm
  • Pressure Transmitter
  • EDU 32 FP, displays volume and calculates flow rate via RS232 and analog-output in realtime (pulse generator required)

V 1.2 / Rev. 2024-07-03 / Subject to alterations.