TG 0.5 Plastic

Technical Drawing

tg000.5 58 v21

Legend

  1. Handle
  2. Support unit for:
    • Thermometer (gas)
    • Thermometer (packing liquid)
    • Manometer
  3. Level
  4. Removable cover
  5. Pulse generator port (option)
  6. Gas outlet
  7. Filling nozzle
  8. Rear plate, detachable
  9. Manometer connection
  10. Gas inlet
  11. Liquid level indicator
  12. Drainage faucet
  13. Levelling foot

Performance Data

Minimum flow Qmin1ltr/h
Standard flow Qstandard50ltr/h
Maximum flow Qmax60ltr/h
Measurement accuracy at standard flow± 0.2%
Measurement accuracy across measuring range± 0.5%
Measuring drum volume0.5ltr/U
Packing liquid quantity, approx2.5ltr
Maximum gas inlet pressure50mbar
Minimum differential pressure (1)0.4mbar
Minimum dial division 0.002ltr
Maximum indication value (2)9,999,999.9ltr
Connection gas in/outlethose barb
Inside-Ø hose barb9.5mm
Outside-Ø hose barb16mm

(1) Differential pressure (= pressure loss) gas inlet gas outlet
(2) Standard Totalizing Roller Counter

Models (materials)

Model Casing Measuring drum Width Height Depth Weight
(without packing liquid)
Max. constant
use temperature
5 PVC-transparent PVC-grey 270 mm 320 mm 350 mm 4.0 kg 40 °C
6 PP-grey PP-grey 270 mm 320 mm 350 mm 3.0 kg 80 °C
7 PVDF PVDF 270 mm 320 mm 350 mm 5.0 kg 80 °C
8 PE-el PE-el 270 mm 320 mm 350 mm 3.0 kg 60 °C

Caution: Before and after measurements with oxygen purge the meter with an inert gas to avoid the danger of explosion! For chemical resistance properties please contact your RITTER distribution partner!

Standard Equipment

  • Multi-Chamber Measuring Drum
  • Magnetic coupling between measuring drum and counter (for an absolutely liquid-tight casing)
  • 8 or 9-digit Totalizing Roller Counter
  • Large One-needle Dial
  • Supports for Thermometer and Manometer
  • Viton sealing
  • Level and Levelling Feet
  • Bubble level for levelling

Built-in Options

Accessories