RITTER MilliGascounter
Type MGC-1 PVC (red)

MilliGascounter

Type MGC-1 PVC (red)

Front View Side View Rear View
(1) Counter unit with LCD display (3) Gas Inlet (5) Air-vent screw for filling
(2) Signal Output (reed contact) (4) Gas Outlet (6) Inspection screw gas channel
(14) Tube adaptors for flexible connecting tubes  

 

Performance Specifications

Minimum flow rate Qmin 1 ml/h Maximum gas inlet pressure 100 mbar
Maximum flow rate Qmax @ measurement accuracy ±3%1)
0.6 ltr/h Minimum gas inlet pressure 5 mbar
Maximum flow rate Qmax @ measurement accuracy ±5%1)  1 ltr/h Gas inlet pressure at mea-surement start, approx. 5) 9 mbar
Volume of measurement chamber 2) 2 ml Gas temperature 6) +10 ..+40 °C
Min. measuring volume (resolution) 3) 2 ml Connection gas in/outlet Hose barb  
Resolution of indication 4) 0.01 ml Diameter hose barb 8o mm
Packing liquid quantity, approx. 120 ml Diameter connecting tube 7i mm

 

1) Because of the physical measurement principle, the measurement error is dependent on the flow rate. The data acquisition software "Rigamo" (accessory) provides an algorithm, which automatically recalculates the actual measurement data to the real values at the respective actual flow rate on the basis of the calibration curve. The remaining error is less than approx. ±1% across the full flow rate range. = approximate value, may vary due to manufacturing tolerances

2) approximate value, can deviate due to production tolerances

3) = volume of measuring chamber

4) Because of calibration factor with 2 decimals

5) Higher gas inlet pressure until gas inlet channel and micro capillary in the base plate are clear of packing liquid

6) For complete MGC unit applies: With temperatures > room temperature (e.g. if placed in heating furnace) a foam formation of the packing liquid was monitored in particular cases.

Materials

  • Casing: PVC (Polyvinyl chloride), red
  • Measurement cell: PVC (Polyvinyl chloride), red
  • Plug-in connectors: Brass, nickel-plated
  • Tube adaptors: PBT (Polybutylenterephthalat)

Standard Equipment

  • Electronic counter / LCD display
  • Display: 6 digits [ml] + 2 decimals
  • Pulse Generator V6.0 as signal output (reed contact, floating output)
  • Twin-chamber measurement cell
  • 2 tube adaptors for flexible connecting tubes
  • 1 cleaning rod for micro capillary per each 1 to 5 MGC's
  • Funnel for filling of liquid
  • 200 ml packing liquid
  • 1.5 m gas connection tubing (PVC)

Accessories

  • Data acquisition software "Rigamo" for Windows
  • Gas connection tubing (PVC)
  • Packing liquid 100 / 500 / 1,000 ml
Subject to alterations

 



Dr.-Ing. Ritter Apparatebau GmbH & Co. KG · Coloniastr. 19-23 · D-44892 Bochum (Germany) · Phone: +49-234-280077 · Fax: +49-234-280078 · E-Mail: mailbox@ritter.de